MODEL NUMBERS
Code SGA-B0990: Fully Automatic Standard Version at 200J
Code SGA-B0991: Fully Automatic Power Version at 360J
CONFIGURATION OPTIOth (BOX CONTENT)
Conf-Norm: Standard Basic Configuration (adult pads, disposable battery)
Conf-Rech: Rechargeable Configuration (adult pads, accumulator, charger station)

 

Technical specifications

DEFIBRILLATORE
Operazione: Versione completamente automatica
Energie: Standard max 200J o Potenza max 360J
Forma d’onda: BTE adattivo (bifasico tronco esponenziale) conforme all’impedenza del torace del paziente
Protocolli: Diversi protocolli di shock per adulti disponibili su richiesta
Default di fabbrica: Adulto Standard crescente 150, 200, 200J
Adulto Potenza crescente 200, 250, 360J

Pediatrico (standard o potenza) 50J fisso

Tempo di carica: ≤9 secondi con una batteria nuova e completamente carica, con una batteria non completamente carica, il tempo di carica sarà più lungo
Tempo di analisi: IEC/EN 60601-2-4 da 4 a 15 secondi
Impedenza: 20-200 ohms
Sensibilità: IEC/EN 60601-2-4 (AHADB, sorgente MITDB), 97%
Specificity: IEC/EN 60601-2-4 (AHADB, sorgente MITDB), 99%
Controls: 2 buttons: ON/OFF,
“i“ info button
Flashing Icons: “connects pads to patient“
“adult/child“ informing on pads type use
“don’t touch patient“ warning to stay clear
“touch patient“ informing it’s safe to touch
Indicators: Status LED indicator informing on device condition Battery gauge with remaining capacity rate
Audible alerts and text display with service alarms
Upgradeable: through a USB cable or memory card or by remote through AMISAVERCLOUD
BATTERY OPTIONS
Type: Li-SOCl2 Disposable, code SAV-C1032
Autonomy: 300 complete rescue cycles (shocks at 200J and CPR)
or 200 complete rescue cycles (shocks at 360J and CPR)
or 35 hours ECG analysis for a new and fully charged battery (*)
Shelf-Life: when stored in original packaging 5 years (*)
Battery-Life: 4 years once installed to AED, assuming one battery insertion test and daily self-test but without switching AED on (*)
Type: Li-ion Accumulator, code SAV-C1033
Recharging time: 2,5 hours with the charger station code SAV-C1034 (*) (recommended to charge every 4 months at least)
Autonomy: 250 shocks at 200J or 160 shocks at 360J or 21 hours in ECG analysis for a new fully charged accumulator (*)
Battery-Life: 2 years or 300 charging cycles (*)
PADS OPTIONS
Type: Disposable, pre-gelled and self-adhesive
Adult: Code SAV-C0846, for patient >8 years or >25 kg
Pediatric: Code SAV-C0016, for patient <8 years or <25 kg
Cable length: 120 cm
Shelf-Life: 30 months
EVENT RECORDING
Internal memory: up to 6 continuous hours of ECG and rescue events
Optional memory: Removable SD card; length of storage depends on card capacity: a 2GB card records up to 100 hours
Data recording: “AED1LOG“ text file with detailed self-test activity
“AEDFILES“ multimedia files with complete recorded information
Event review: “Saver View Express“ data manager software
PHYSICAL
Size: 29,5 x 23 x 11,5 cm
Weight +/- 2,65 Kg
ENVIROMENTAL
Operating temperature: 0°C to 55°C (32°F TO 131°F)
Storing/Shipping temperature: -40°C to 70°C (-40°F TO 158°F)
Humidity: 10% to 95% relative humidity non condensing
Sealing (IP Protection): IEC/EN 60529 class IP56; splash proof, dust protected
Shock/Drop Abuse Endurance: IEC/EN 60601-1 clause 21; 1 meter drop, impact, force, rough handling, mobile tolerance
Electrostatic Discharge: IEC/EN 61000-4-2
Electromagnetic Compatibility: IEC/EN 60601-1-2 Emission, Immunity
Electrical Protection: IEC/EN 60601-1; Internally Powered, Type BF
Directive 93/42/CEE and 2007/47/CE: Class IIb
Radio Equipment Directive (RED): Directive 2014/53/UE
GEOLOC MODULE
Frequency: GSM: 850, 900, 1800, 1900 MHz;
UMTS: 900, 2100 MHz
GPS: 1575, 1600 MHz
GEOLOC BATTERY OPTIONS
Type: Li-SOCl2 Disposable, code SAV-C1038
Shelf-Life: when stored in original packaging 5 years (*)
Battery-Life: 4 years once installed to AED, assuming one battery insertion test and daily self-test but without switching AED on (*)
Type: Li-ion Accumulator, code SAV-C1039
Recharging time: 2,5 hours with the charger station code SAV-C1040 (*)
(recommended to charge every 4 months at least)
Battery-Life: 2 years or 300 charging cycles (*)
(*)Temperature at 20°C Humidity 45% non-condensing

Geo Saver Fully Automatic - Technical Data Sheet

Download